Detector Geiger-Muller LND 7121 (ZP1200, ZP1201)

    SEAP
    ID
    DA13776563
    Data
    05 Octombrie 2017
    Valoare
    0 RON
    Stare
    Conditii refuzate
    Autoritatea contractantaLocalitate
    Bucuresti, Bucuresti
    FurnizorTipul contractului
    -
    Cod CPVDescriere:
    Detector Geiger-Muller LND 7121 (ZP1200, ZP1201)
    Achizitii
    0 RON
    Cantitate: 30
    Unitate masura: bucata
    Detector Geiger-Muller LND 7121 (ZP1200, ZP1201)
    7121 Gamma Detector WINDOW SPECIFICATIONS Effective diameter (inch/mm) 0.36/9.1 Material Stainless Steel Thickness (inch/mm) 0.010/0.254 GENERAL SPECIFICATIONS Gas filling Ne +Halogen Cathode material 446 Stainless Steel Maximum length (inch/mm) 1.94/49.2 Effective length (inch/mm) 1.5/38.1 Maximum diameter (inch/mm) 0.59/15.1 Connector Pin Operating temperature range °C -10 to +75 ELECTRICAL SPECIFICATIONS Recommended anode resistor (meg ohm) 10 Maximum starting voltage (volts) 325 Recommended operating voltage (volts) 500 Operating voltage range (volts) 450 - 650 Maximum plateau slope (%/100 volts) 6 Minimum dead time (micro sec) 90 Gamma sensitivity Co60 (cps/mr/hr) 18 Tube capacitance (pf) 3 Weight (grams) 8 Maximum background shielded 50mm Pb+3mm(cpm) 10