Verificare MasterScreen Body-Diff

    SEAP
    ID
    DA34076276
    Data
    22 Septembrie 2023
    Valoare
    1.150 RON
    Stare
    Oferta acceptata
    Autoritatea contractantaLocalitate
    Bucuresti, Bucuresti
    FurnizorTipul contractului
    Servicii
    Cod CPVDescriere:
    Verificare aparat Jaeger MasterScreen Body-Diff. Se verifica: - ambient module - pneumotach O-rings - zero and gain at the PT-pressure transducer (use EASI-Test and reference PT-screen) - zero (and gain*) at the PV-pressure transducer (use EASI-Test and reference pressure*) - shutter valves for easy rotation (torque) - shutter valve for leak - PT-flange, fittings and plastic elbows for cracks (replace if necessary) - zero and gain at PB- and PV-pressure transducer (use EASI-TEST and Body Calibration) - flexible tube for wear and leak (replace if necessary) - check temperature insulation at Helium sensor (p/n 845 568) - analyzer (sample) pump alignment and adjust if necessary - zero and gain factor at the CO analyzer fuel cell (use EASI-Test and gas calibration; analyzer fuel cells with gain factor >1.6 should be replaced) - support arm tension (brake screw) - 50ml sine pump - Body Close Valve hub - Pressure Safety Valve (if available) for leak and proper function (o-ring 851032) - all 3 door lock brackets (replace if necessary) - body box for leak (use body box calibration) Costul componentelor care necesita schimbare si manopera nu este inclus. Nu toate componentele de service pot fi mentinute in permanenta in stocul distribuitorului; componente de service care nu fac parte din procedura de intretinere preventiva si nu se afla in stoc vor fi comandate la producator, in cazul in care este necesara schimbarea lor; timpul standard de livrare este de 2-4 saptamani valabil
    Achizitii
    1.150 RON
    Cantitate: 1
    Unitate masura: bucata
    Verificare MasterScreen Body-Diff
    Verificare aparat Jaeger MasterScreen Body-Diff. Se verifica: - ambient module - pneumotach O-rings - zero and gain at the PT-pressure transducer (use EASI-Test and reference PT-screen) - zero (and gain*) at the PV-pressure transducer (use EASI-Test and reference pressure*) - shutter valves for easy rotation (torque) - shutter valve for leak - PT-flange, fittings and plastic elbows for cracks (replace if necessary) - zero and gain at PB- and PV-pressure transducer (use EASI-TEST and Body Calibration) - flexible tube for wear and leak (replace if necessary) - check temperature insulation at Helium sensor (p/n 845 568) - analyzer (sample) pump alignment and adjust if necessary - zero and gain factor at the CO analyzer fuel cell (use EASI-Test and gas calibration; analyzer fuel cells with gain factor >1.6 should be replaced) - support arm tension (brake screw) - 50ml sine pump - Body Close Valve hub - Pressure Safety Valve (if available) for leak and proper function (o-ring 851032) - all 3 door lock brackets (replace if necessary) - body box for leak (use body box calibration) Costul componentelor care necesita schimbare si manopera nu este inclus. Nu toate componentele de service pot fi mentinute in permanenta in stocul distribuitorului; componente de service care nu fac parte din procedura de intretinere preventiva si nu se afla in stoc vor fi comandate la producator, in cazul in care este necesara schimbarea lor; timpul standard de livrare este de 2-4 saptamani valabilitatea ofertei 60 de zile