Pachet etalonari CERTELAB
SEAPIDStare
DA31527854
Data03 Octombrie 2022
Valoare26.350 RON
Oferta refuzata
Autoritatea contractantaLocalitateSuceava, Suceava
FurnizorTipul contractuluiServicii
Cod CPVDescriere:Etalonare in regim acreditat ISO17025 pentru R&S ENV216, Etalonare in regim acreditat ISO17025 pentru SMR 20, Etalonare in regim acreditat ISO17025 pentru ETS HI-6053
Achizitii7.350 RON
Cantitate: 1
Unitate masura: bucata
Etalonare in regim acreditat ISO17025 pentru R&S ENV216
Etalonare in regim acreditat ISO17025 pentru R&S ENV216.
Accredited calibration in accordance with the European Standard EN 55016-1-2, ed 2, 2014.
Frequency range 9 kHz – 30 MHz.
Measured parameters:
- Voltage transmission (between respective EUT Socket's terminal and RF test receiver output, Mains plug terminated using 50 Ω resistors)
- Impedance at EUT (Mains plug terminated by Open / Short)
- Impedance at RF connector (Mains plug terminated by Open / Short)
- Isolation (transmission between respective Mains Socket and RF test receiver output, EUT socket terminated using 50 Ω resistors)8.750 RON
Cantitate: 1
Unitate masura: bucata
Etalonare in regim acreditat ISO17025 pentru SMR 20
Etalonare in regim acreditat ISO17025 pentru SMR 20:
- Reference oscillator frequency check
- Frequency flatness, L = {3, -10} dBm
- Linearity of output level, f = 1 GHz
- Harmonic distortion (2nd, 3rd) for f up to 13 GHz / 8 GHz
- Output frequency check
- Maximum output level check10.250 RON
Cantitate: 1
Unitate masura: bucata
Etalonare in regim acreditat ISO17025 pentru ETS HI-6053
Etalonare in regim acreditat ISO17025 pentru ETS HI-6053:
1. Frequency response (E = 20 V/m):
f = {10, 27, 50, 100, 200, 300, 400, 500, 600, 700, 800, 900, 1000, 1500, 1800, 2450, 2700, 3000, 3500, 4000, 5000, 6000, 7000, 8000, 9000, 10000} MHz.
Expected uncertainty of calibration is approx. 0.5–1.5 dB (a higher uncertainty can be expected especially at the 7 GHz frequency point)
2. Linearity check (f = 100 MHz): E = {3, 10, 20, 30, 50, 100, 200, 300} V/m
Expected uncertainty of calibration is approx. 0.5–1.0 dB
3. Isotropy check (f = 100 MHz)
Expected uncertainty of calibration is approx. 0.7 dB